Reverse-phase HPLC
A sample is pushed through a column packed with a non-polar stationary phase. Compounds that bind the phase more strongly take longer to come off, so the mixture separates in time. A detector, usually UV at 214 nm where the peptide bond absorbs, records what elutes and when. The output is a chromatogram: a baseline with peaks on it.
Purity is computed as main peak area ÷ total peak area, expressed as a percentage. That is the whole definition, and its limits follow directly from it:
- It is a proportion of what the detector saw. Anything that does not absorb at the chosen wavelength is invisible to it, including most salts and the counter-ion.
- An impurity that co-elutes with the main peak is counted as the main peak. Purity is only as good as the separation.
- It says nothing about how much peptide is in the vial. A vial containing 3 mg of peptide and a vial containing 8 mg can both report 99%.
A gradient, column and wavelength should accompany the number. A purity figure with no method behind it is not reproducible.