What each test tells you, and what it can’t
Each measurement behind a certificate answers a different question and has a limit. Knowing both lets you read the certificate, not just the headline number.
01 · Reverse-phase HPLC
Purity
What fraction of the detected material is the target compound?
The sample is pushed through a hydrophobic column, and a detector records each component as a peak. The target peak's share of the total peak area is the purity figure on the certificate.
What it can’t tell you: It separates by hydrophobicity, not identity. A closely related impurity can sit inside the main peak, and the figure doesn't confirm the peak is the compound you ordered.
02 · Mass spectrometry
Identity
Is this the compound it's labeled as?
The instrument measures mass-to-charge ratio, and the observed mass is compared with the mass calculated from the expected sequence.
What it can’t tell you: Mass confirms composition, not sequence order. Two peptides with the same residues in a different order weigh the same.
03 · Net peptide content
Content
How much of the vial's mass is actually peptide?
Lyophilized peptides are usually supplied as a salt and hold some moisture, so this figure separates peptide from counter-ion and bound water.
What it can’t tell you: It applies to the tested lot only, and it's the figure to use when an exact amount matters more than the labeled mass.
04 · Appearance and solubility
Condition
Does the material look and behave as expected?
The lab records what the material looks like. A white to off-white, intact cake is the usual expected condition.
What it can’t tell you: It's a basic check. It says nothing about purity or identity on its own.